Register NowReference Interval
0.00–0.18 pg/mL
Overview:
Use: This test is used to aid in the biological identification of Alzheimer's disease by measuring phosphorylated tau 217, which has been shown in the literature to be a surrogate of amyloid pathology. Additionally, pTau-217 can be used for monitoring patients on newly approved anti-amyloid therapy.
Clinical cutoff value was established using samples from a well-established patient cohort characterized with amyloid PET data. A pTau-217 value of >0.18 is a reported surrogate marker for beta-amyloid pathology and can be used to facilitate biological identification of Alzheimer's disease.1 pTau-217 has also been used in clinical trials to monitor patients on anti-amyloid therapy.2,3
Limitations: This test is performed by Fujirebio Lumipulse chemiluminescent enzyme immunoassay (CLEIA). Values obtained with different methods cannot be used interchangeably. The validated limit of quantification is 0.06 pg/mL. Assay detection limit is 0.03 pg/mL.
This test was developed and its performance characteristics determined by Labcorp. It has not been cleared or approved by the Food and Drug Administration.
Methodology: Fujirebio Lumipulse chemiluminescent enzyme immunoassay (CLEIA)
Reference Interval
0.00–0.18 pg/mL
Collection Instructions:
Specimen: Plasma, frozen
Volume:1 mL
Minimum Volume: 0.75 mL (Note: This volume does not allow for repeat testing.)
Container: Lavender-top (EDTA) tube
Collection: Draw blood in lavender-top (EDTA) tube. Invert to mix with preservatives. Centrifuge and transfer plasma to a labeled plastic transport tube and freeze.
Storage Instructions: Freeze